Dosimag
Installation
Hauser
21
Linear filling system
1
3
4
2
A0003762
1
Measuring device
2
Tank
3
Batching valve
4
Vessel
6.2
Mounting the measuring device
6.2.1
Required tools
For sensor
For flanges and other process connections:
• Screws, nuts, seals etc. are not included in the scope of supply and must be provided by
the customer.
• Appropriate mounting tools
6.2.2
Preparing the measuring device
1. Remove all remaining transport packaging.
2. Remove any protective covers or protective caps present from the sensor.
3. Remove stick-on label on the electronics compartment cover.
6.2.3
Mounting the measuring device
L
WARNING
Danger due to improper process sealing!
‣
Ensure that the inside diameters of the gaskets are greater than or equal to that of the
process connections and piping.
‣
Ensure that the gaskets are clean and undamaged.
‣
Install the gaskets correctly.
Depending on the order option, the measuring device is supplied with or without pre-
installed process connections. Pre-installed process connections are secured to the
measuring device using 4 hexagonal-headed bolts.
‣
Ensure that the direction of the arrow on the nameplate of the sensor matches the flow
direction of the fluid.
Depending on the application and pipe length, the measuring device may need to be
supported or additionally secured.