Orientation
The direction of the arrow on the sensor nameplate helps you to install the sensor
according to the flow direction (direction of medium flow through the piping).
An optimum orientation can prevent buildup and the accumulation of gas or air in the
measuring tube.
Valves
Never install the sensor downstream from a filling valve. If the sensor is completely empty
this corrupts the measured value.
Correct measurement is only possible if the pipe is completely full. Perform sample
fillings before commencing filling in production.
1
2
2
1
3
3
A0003768
1
Measuring device
2
Filling valve
3
Container
Filling systems
The pipe system must be completely full to ensure optimum measurement.