Post-installation check
Is the device undamaged (visual inspection)?
Does the measuring device conform to the measuring point specifications?
For example:
• Process temperature
• Process pressure→ 65
• Ambient temperature → 63
• Measuring range
Horizontal position of the measuring electrode plane?
Has the correct orientation for the sensor been selected ?
• According to sensor type
• According to medium temperature
• According to medium properties (outgassing, with entrained solids)
Does the arrow on the sensor nameplate match the direction of flow of the fluid through the
piping ?
Are the measuring point identification and labeling correct (visual inspection)?
Is the measuring device adequately protected against vibration (attachment, support)?→ 19
Are the inlet and outlet runs to respected?→ 18