Technical data
Dosimag
64
Hauser
• Protect the measuring device against direct sunlight during storage in order to avoid
unacceptably high surface temperatures.
• Select a storage location where moisture cannot collect in the measuring device as
fungus or bacteria infestation can damage the liner.
• If protection caps or protective covers are mounted these should never be removed
before installing the measuring device.
Degree of protection
As standard: IP67, type 4X enclosure
Shock resistance
Acceleration up to 2 g based on IEC 60068-2-6
Vibration resistance
Acceleration up to 2 g based on IEC 60068-2-6
Interior cleaning
• Cleaning in place (CIP)
• Sterilization in place (SIP)
Observe the maximum medium temperatures → 64
Electromagnetic
compatibility (EMC)
According to IEC/EN 61326
For details, refer to the Declaration of Conformity.
16.9 Process
Medium temperature range
Sensor
–20 to +130 °C (–4 to +266 °F)
Cleaning
+150 °C (+302 °F) / 60 min for CIP and SIP processes
Seals
• EPDM: –20 to +130 °C (–4 to +266 °F) (max. +150 °C (302 °F) for cleaning
• Silicon:–20 to +130 °C (–4 to +266 °F)
• Viton: 0 to +150 °C (+32 to +302 °F)