Dosimag
Commissioning
Hauser
39
Parameter
Prerequsite
Description
Selection / User
entry
Factory setting
Failure mode
–
Define output behavior in
alarm condition.
• Actual status
• Open
• Closed
Open
Invert output signal
–
Invert the output signal.
• No
• Yes
• Pulse/freq./switch
output 1: yes
• Pulse/freq./switch
output 2: no
10.4.4 Low flow cut off
The
Low flow cut off
submenu contains the parameters that must be set in order to
configure the low flow cut off.
Navigation
"Setup" menu → Low flow cut off
‣
Low flow cut off
Assign process variable
On value low flow cutoff
Off value low flow cutoff
Pressure shock suppression
Parameter overview with brief description
Parameter
Prerequsite
Description
Selection / User
entry
Factory setting
Assign process variable
–
Select process variable for low
flow cut off.
• Off
• Volume flow
Volume flow
On value low flow cutoff
In the
Assign process variable
Volume flow
option is
selected.
Enter on value for low flow cut
off.
Signed floating-point
number
Depends on country
and nominal
diameter
Off value low flow cutoff
In the
Assign process variable
Volume flow
option is
selected.
Enter off value for low flow cut
off.
0 to 100.0 %
50 %
Pressure shock suppression
In the
Assign process variable
Volume flow
option is
selected.
Enter time frame for signal
suppression (= active pressure
shock suppression).
0 to 100 s
0 s
10.5 Advanced settings
The
Advanced setup
submenu together with its submenus contains parameters for
specific settings.