Dosimag
Technical data
Hauser
63
Accuracy of outputs
The outputs have the following base accuracy specifications.
Pulse/frequency output
o.r. = of reading
Accuracy
Max. ±50 ppm o.r. (across the entire ambient temperature range)
Repeatability
o.r. = of reading
DN 25 (500 ml/s), DN 15 (200 ml/s), DN 8 (50 ml/s), DN 4 (10 ml/s); 400 μS/cm
Batch time t
a
[s]
Relative standard deviation in relation to the batched volume [%]
1.5 s < t
a
< 3 s
0.4
3 s < t
a
< 5 s
0.2
5 s < t
a
0.1
DN 15K (200 ml/s); 400 μS/cm
Batch time t
a
[s]
Relative standard deviation in relation to the batched volume [%]
1.5 s < t
a
< 3 s
0.25
3 s < t
a
< 5 s
0.12
5 s < t
a
0.08
Influence of ambient
temperature
Pulse/frequency output
Temperature coefficient
No additional effect. Included in accuracy.
16.7 Installation
"Mounting requirements"
16.8 Environment
Ambient temperature
range
Temperature tables
Observe the interdependencies between the permitted ambient and fluid
temperatures when operating the device in hazardous areas.
For detailed information on the temperature tables, see the separate document
entitled "Safety Instructions" (XA) for the device.
Storage temperature
The storage temperature corresponds to the ambient temperature range of the transmitter
and sensor.