Diagnostics and troubleshooting
Dosimag
48
Hauser
Diagnostic
number
Short text
Remedy instructions
Status signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
Diagnostic of sensor
004
Sensor
1. Change sensor
2. Contact service
S
Alarm
062
Sensor connection
1. Check sensor connections
2. Contact service
F
Alarm
082
Data storage
1. Check module connections
2. Contact service
F
Alarm
083
Memory content
1. Restart device
2. Contact service
F
Alarm
Diagnostic of electronic
242
Software incompatible
1. Check software
2. Flash or change main
electronics module
F
Alarm
270
Main electronic failure
Change main electronic module
F
Alarm
271
Main electronic failure
1. Restart device
2. Change main electronic
module
F
Alarm
272
Main electronic failure
1. Restart device
2. Contact service
F
Alarm
273
Main electronic failure
Change electronic
F
Alarm
281
Electronic initialization
Firmware update active, please
wait!
F
Alarm
311
Electronic failure
1. Reset device
2. Contact service
F
Alarm
322
Electronic drift
1. Perform verification manually
2. Change electronic
S
Warning
Diagnostic of configuration
410
Data transfer
1. Check connection
2. Retry data transfer
F
Alarm
411
Up-/download active
Up-/download active, please wait C
Warning
438
Dataset
1. Check data set file
2. Check device configuration
3. Up- and download new
configuration
M
Warning
442
Frequency output 1 to 2 1. Check process
2. Check frequency output
settings
S
Warning
443
Pulse output 1 to 2
1. Check process
2. Check pulse output settings
S
Warning
453
Flow override
Deactivate flow override
C
Warning
484
Simulation failure mode Deactivate simulation
C
Alarm
485
Simulation measured
variable
Deactivate simulation
C
Warning
500
Electrode difference
voltage too high
1. Check process cond.
2. Increase system pressure
F
Alarm
Diagnostic of process
832
Electronic temperature
too high
Reduce ambient temperature
S
Warning
833
Electronic temperature
too low
Increase ambient temperature
S
Warning