EM TEST
UCS 500 Series
Operating Manual
V 5.40
63 / 99
11.2.2.
Standard Test Routines
Standard test routines
F1 : IEC 61000-4-11 (ac power supply mains)
F2 : IEC 61000-4-29 (dc power supply systems)
F3 : EN 61000-6-1
Generic standard
F4 : EN 61000-6-2
Generic standard
F7 : Manual test routine
F1
F2
F3
F4
F5
F6
F7
11.2.2.1.
F1: IEC 61000-4-11 (AC power supply mains)
Page 3
Power Fail
IEC 61000-4-11
As long as the external variac MV2616
is used, controlled by an analogue 0-
10V control voltage, the test is
conducted automatically. If this option is
not available the manual test routine
shall be used.
F1:
Level 1
0%
10.0 ms
F2:
Level 2
0%
20.0 ms
F3:
Level 3
40%
200 ms
F4:
Level 4
70%
500 ms
F5:
Level 5
80% 5000 ms
F6:
Short Interruptions
F7:
Voltage Variation
F1
F2
F3
F4
F5
F6
F7
Using a tapped transformer V 4780, the user must connect the PF2 UCS input with the correct plug at the V4780.
At 0% selection the PF2 input must be connected to the neutral (0V) plug.
F1 ... F5: DIPS Level 1 to Level 5
Page 4 ( Show parameter)
Power Fail
IEC 61000-4-11
Level 3
V2% =
40%
A
= 0 dgr
td
=
200ms
tr
= 10s
n
= 3.
Counter
START / STOP
000002
F1
F2
F3
F4
F5
F6
F7
F6: Short interruptions
Page 4 ( Show parameter)
Power Fail IEC 61000-4-11
Interruption
V2% =
0 %
A
= 0 dgr
td
=
5000ms
tr
= 10s
n
= 3
Counter
START / STOP
000002
F1
F2
F3
F4
F5
F6
F7