4.5.
Setup
Firmware V 6.
01
ax (UCS 500N5) V9.
04
ax (N7x)
SETUP
SETUP
F1 : Change language / Sprache ändern
F1 : Change language / Sprache ändern
F2 : LCD backlighting
F2 : LCD backlighting
F3 : Interfaces
F3 : Interfaces
F4 : Beep
F4 : Beep
F5 : Timer
F5 : Timer
F6 : Set voltage
F6 : Set voltage
low / high
F7 : Magnetic field correction factors
F7 : Magnetic field correction factors
F1
F2
F3
F4
F5
F6
F7
F1
F2
F3
F4
F5
F6
F7
F1 Change language
The user can chose between two languages, German and English.
F2 LCD backlighting
With the use of F2 the backlighting can be switched on or off.
Additionally the
Auto Off
function can be programmed to switch off the backlighting after a defined time that the
equipment has not been in operation (1 - 30minBecause of the limited lifetime of LCD displays, approx. 10,000h
this function should always be activated.
F3 Interfaces
This menu will help the user to define the status of the integrated serial and parallel interfaces, e.g. the baud rate
of the RS 232 or the address of the IEEE interface.
F4 Beep
F1 is the selector for the keyboard beeper ON/OFF mode. (Short beep at every keyboard hit)
F3 is the selector for the countdown-beeper ON/OFF mode. (Short beep before surge pulse releasing )
To indicate that a running test is finished the beeper sounds always 3 times (not changeable).
F5 Operating time
Pressing of F5 will show the different operating time and status of the test equipment.
Operating time:
Total time where the UCS is powered on.
Testing Time:
Total time during a running test.
DI
– Status:
Service information about internal digital inputs.
SCR
– Status:
Service information about surge switch operation
F6 Set voltage low/high
Firmware V 6.01ax (UCS 500N5) V9.04ax (N7x)
This Firmware version offers
two separate voltage settings
for user that needs to test equipment for 115 V and
230V mains supply. Thanks to the two settings, the user can keep the Motorvariac in the 230V position
Using a MV 2616 with voltage selector switch:
- Set the voltage selector to the 230V position for low and high mains voltage supply.
- Use the Firmware with low and high settings and use the MV 2616 for any supply in the 230 V position.
The operator can specify the following parameters:
F1: Setup Main Voltage to low 115 V
.
Motorvariac parameter settings for lower main voltage. (115 V)
F2: Setup Main Voltage to high 230 V
Motorvariac parameters for higher main voltage. (230 V)
F3: Motorvariac high (low)
Selection and indicate
of the active “Main Voltage” setting. The
selected setting is used in the power Fail tests for motorvariac
control
The procedure for low and high parameters voltage setting is the same as described in F6 Set Voltage below.
F6 Set voltage
For control of an external power supply source an analogue control voltage can be programmed (0-10V dc).
The operator can specify the following parameters:
F1: Max. variac voltage
V [ V ] Maximal output voltage at 100% position of the external motor variac or dc
controlled voltage source. The analogue reference value for the max. voltage is 10 V.
F2: Mains supply voltage
V [ V ] of the device under test. This voltage shall be specified by the operator and
depends on the type of equipment under test. The variac normally is automatically set to this output voltage.
F3: Default channel
. The operator select the default channel PF1 or PF2 from which the EUT is powered.