![EMTEST UCS 500 N7 Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/emtest/ucs-500-n7/ucs-500-n7_manual_2404983016.webp)
EM TEST
UCS 500 Series
Operating Manual
V 5.40
16 / 99
4.3.
Main Menu
MAIN MENU
F1 : Burst
IEC 61000-4-4
F2 : Surge
IEC 61000-4-5/9
F3 : Power Fail
IEC 61000-4-11/8/29
F4 : Ringwave
IEC 61000-12
F5 : TSurge
F7 : Service
F1
F2
F3
F4
F5
F6
F7
F1 Burst test
With function key F1 the user can select Burst Test as per
IEC 61000-4-4
. The test pulses are fast transients with
a pulse shape of 5/50ns.
Attention:
The generator covers the new specifications given in the draft revision IEC 61000-4-4 edition 2. This means a new spike
frequency which is selectable between 5 kHz and 100 kHz. The burst duration is automatically matched between 15ms
and 0.75ms. The common mode coupling is new with all couplings at the same time.
F2 Surge test
With function key F2 the user can select Surge Test as per
IEC 61000-4-5
. The test pulses are high energy pulse
with a voltage shape of 1,2/50
s and a short circuit current shape of 8/20
s.
In addition the test procedure for Pulse Magnetic Field testing as per
IEC 61000-4-9
is included in this menu.
F3 Power Fail test
With function key F3 the user can select the Power Fail Test as per
IEC 61000-4-11
.
The simulator will generate voltage dips and voltage variations with preselectable parameters.
IEC 61000-4-11
is
valid for ac power supply systems. For dc supply system in future
IEC 61000-4-29
shall be recommended.
In addition the test procedure for 50/60Hz Magnetic Field Testing as per IEC 61000-4-8 is included within the
menu.
The Power fail menu includes a
Voltage Variation
test as per IEC 61000-4-11 as test procedure. The simulator
will generate voltage variations in the second range.
Attention:
The generator covers already the new specifications given in FDIS IEC 61000-4-11: 2004. This means the introduction of
a new 80% test level as well as a new function for the Voltage Variation test.
F4 Ringwave as per IEC 61000-4-12
(UCS500 N7)
With function key F4 the user can select the Ring Wave Test as per
IEC 61000-4-12
or
ANSI/IEEE.
The test
pulses are single-shot with an oscillating and decaying waveshape.
F5 Telecom Surge Test
With the Option TSurge a Telecom Surge pulse 10/700
s as per IEC 61000-4-5, ITU, ETS or FCC part 68 Pulse
B is available.
F7 Service
Setup and servicing routines are available.