Voltage Dips as per IEC 61000-4-11
Dips Module for voltage Dips and short interruptions.
11.1. Test setup for DIPS and Interruption tests
Voltage DIPS
For voltage DIPS a connection must be made from the
PF2 UCS input to the motorvariac PF2 or V4780, where
the reduced voltage is present.
Interruption tests
For Interrupts connect the PF2 UCS input to the neutral
as follows:
-
Motorvariac
: PF2 output (default setting) set
motorvariac to zero volt.
-
V 4780
:
Connect the cable to the N output at
the V 4780
11.2. Operation
The Surge menu offers different test routines for burst testing. The bottom line shows the selected nominal
voltage (Change in the menu F3 Setup / F6 Set voltage…)
Power Fail
IEC 61000-4-11 / 8 / 29
F1 : Quick Start
F2 : Standard test routines
F3 : User test routines
F4 : Magnetic field
Motorvariac Mains Voltage 115V
Display Firmware V9.
04
ax and higher
F1
F2
F3
F4
F5
F6
F7
F1 Quick Start
Easy and fast online-operation of the equipment.
F2 Standard test routines
The operator can select between various preprogrammed test routines as required in different standards.
With the function key F4 the operator can select the standard routine for 50/60Hz magnetic field testing as per
IEC 61000-4-8.
F3 User test routines
The operator can select between various preprogrammed test routines which helps to accelerate testing and
which are very helpful especially during design.
F4 Magnetic Field
The operator can perform 50 / 60 Hz magnetic field test as per IEC / EN 61000-4-8.