EM TEST
UCS 500 Series
Operating Manual
V 5.40
41 / 99
9.1.1.
Quick Start
Easy and very fast operation of all standard functions of the equipment. The latest simulator settings are stored
automatically and will be recalled when Quick Start is next selected.
Burst
Quick start
V
=
500V
f
=
5kHz
td
=
15ms
tr
=
300ms
cpl =
L
+/-
=
+
T
=
01:00min
START CHANGE
F1
F2
F3
F4
F5
F6
F7
Press
CHANGE
and the test parameters parameter can be changed.
Select the desired parameter with the related function key and change the value by turning the front panel knob.
The cursor allows the user to define the digit to be changed (fast or slow change).
Press
START
and the test starts immediately with the displayed test parameters.
The operator now can navigate with the
Cursor
from parameter to parameter. The blinking parameter can be
changed by turning the front panel knob.
Burst specification as per IEC 61000-4-4 Ed2 : 2004
f
= 5kHz
td = 15ms
tr = 300ms
f
= 100kHz
td = 0.75ms
tr = 300ms