EM TEST
UCS 500 Series
Operating Manual
V 5.40
78 / 99
12.
Ringwave Immunity as per IEC 61000-4-12
Ringwave Module 0.5
s
– 100kHz
12.1. Operation
The menu offers different test routines for ring wave testing.
Ring wave
F1 : Quick Start
F1
F2
F3
F4
F5
F6
F7
F1 Quick Start
Easy and fast online-operation of the equipment.
12.1.1.
Quick Start
Easy and very fast operation of all standard functions of the equipment. The latest simulator settings are stored
automatically and will be recalled when Quick Start is next selected.
Ring wave
Quick start
V
= 2000 V
Imp =
12 Ohm
A
= 0 dgr
+/-
=
+
cpl
= L+N
–PE
tr
=
1s
tri
= Auto
n
=
endl
START CHANGE
F1
F2
F3
F4
F5
F6
F7
Press
START
and the test routines begin to work.
Press
CHANGE
and the actual parameters can be changed.
All function keys except F2 (manual trigger) can
Stop
the test routine.
Page 3
(START)
Ring wave
Quick start
V
= 2000 V
Imp =
12 Ohm
A
= 0 dgr
+/-
=
+
cpl
= L+N
–PE
tr
=
1s
tri
= Auto
n
=
endl
USet = 2000V
U = + 2000V Counter
STOP
I = + 0A
0000043
F1
F2
F3
F4
F5
F6
F7
The user can select the parameter with the related function key and can change the value with the knob. The
cursor allows the user to define the value of the digit which should be changed (fast or slow change).
Pressing the ESC button will bring the user back to the previous level from where the test can be restarted with
new levels. After restart the actual test time is displayed. All functions keys except F2 (MAN TRIGGER) can stop
the test routine. The latest setting will be displayed.
Any pressing of a function key will indicate the functions START, CHANGE or CONTINUE. F3 will continue the
same test routine. Also the test time will continue running. If the user selects at first START or CHANGE the test
will be stopped completely.