EM TEST
UCS 500 Series
Operating Manual
V 5.40
38 / 99
8.
Delivery Groups
8.1.
Basic equipment
Generator type UCS 500N5 with recommended modules ( Burst, Surge, Power Fail, TSurge)
UCS 500N7 with recommended modules ( Burst, Surge, Power Fail, TSurge, Ringwave)
Mains cable
Mains cable for the EUT supply
Cable black 0.5m 4mm Safety plug for phase synchronization (PF1-SYNC)
K - USB cable
USB A / B , 3m
Adapter for power cable
Manual (on memory stick)
Calibration certificate
8.2.
Accessories and options
ESA Adapter for adapt different power mains connectors
As accessories adapters to different power mains connectors are part of the
delivery of EM TEST surge generators. E.g. these are adaptors for Schuko - US
- AUS - UK power mains connectors.
Most of these commercially available power mains sockets cannot withstand
surge voltages higher than 4000V. Therefore each of these EM TEST adaptors
are labeled to be
used up to 4000V
maximum.
In case that the EM TEST surge generator can generate higher surge voltages
than 4000V a sparkover at the power mains socket may occur.
Customers shall be aware of this matter and shall
not use higher voltages than
labeled
on the adapter.
For
testing higher voltages
it is necessary to use high voltage cables with
sufficient isolation and safety banana connectors connected direct to the
generator output plugs (L-N-PE).
Burst
CA EFT Kit
Calibration Adapter for Burst verification on EUT KW50 +
KW1000
KW 50
Matching resistor 50
with integrated attenuator (1:100)
KW 1000
Matching resistor 1000
with integrated attenuator (1:500)
HFK
Capacitive coupling clamp as per IEC 61000-4-4 to couple the fast transients
to signal and data lines.
CA HFK
Calibration kit for capacitive coupling clamp as per IEC 61000-4-4 Ed.3
- Flexible plate insulated
- Support for KW 50 and adapter to KW 50
CNE 503
external 3 phase coupling/decoupling network
- EUT Mains
400 V rms max. // 480V for USA
- Rated current
In = 16 A / 32A / 63A / 100 A rms
- Mains frequency
50/60 Hz
- Coupling to all phases
Lx, N, PE. Control with EFT 500.
ITP Radiating set for immunity and emission
A6dB
as additional 6 dB / 50
attenuator of the test signal
PUW 500
EUT monitoring