EM TEST
UCS 500 Series
Operating Manual
V 5.40
8 / 99
Special models
Special models have the index UCS500Nx
S
x. The difference to the standard models are: Voltage- and current
ranges. The operation is t
he same as by the standard UCS equipment’s.
Model
Pulse voltage
CDN
UCS 500 M4 S1
4kV
250V
32A single phase
UCS 500 M4 S2
4kV
690V
16A single phase
UCS 500 M4 S3
4kV
690V
32A single phase
UCS 500 M4 S5
4kV
280V
16A single phase
UCS 500 M6 S2
6kV
690V
16A single phase
ANSI
Model
Pulse
EUT
Modules
UCS 500 M6B S1 6kV
250V 32A
EFT/6, VCS/6 and PFS/6; 1ph CDN 250V / 32A (ANSI)
module RWG/6 not available
UCS 500 M6B S2 6kV
690V 16A
EFT/6, VCS/6 and PFS/6; 1ph CDN 690V / 16A (ANSI) 480V dc
module RWG/6 not available
1.2.
Standards covered by UCS N Series
A fully equipped UCS 500N covers the following standards
UCS 500
N5
N7
- IEC 61000-4-4
Burst
X
X
- IEC 61000-4-5
Surge
X
X
- IEC 61000-4-8
50/60Hz Magnetic Field
X
X
- IEC 61000-4-9
Pulse Magnetic Field
X
X
- IEC 61000-4-11
Voltage Dips
Voltage Interruptions
Voltage Variations for ac power mains supply
X
X
- IEC 61000-4-12
Ringwave
X
- IEC 61000-4-29
Voltage Interruptions
Voltage Dips for dc power supply systems
X
X
- ANSI
Surge with 2
coupling
X
- ITU and ETS
Telecom Surge
(X)
(X)
- FCC
Part 68
(X)
(X)