EM TEST
UCS 500 Series
Operating Manual
V 5.40
24 / 99
6.
Technical data
6.1.
EFT Electrical Fast Transients Burst as per IEC 61000-4-4
Test Level
Model N5
Model N7
Open circuit *
200V - 5500V ± 10% Step 20V
200V
– 5500V ± 10% Step 50V
Wave shape into a 50
load
100V
– 2750V
100V
– 2750V
Rise time tr
5ns ± 30%
5ns ± 30%
Pulse duration td
50ns
30%
50ns
30%
Wave shape into a 1000
load
200V
– 5500V
200V
– 5500V
Rise time tr
5ns ± 30%
5ns ± 30%
Pulse duration td
35ns - 150ns
35ns
– 150ns
Source impedance
Zq = 50
± 20%
Zq = 50
± 20%
Polarity
Positive / negative
positive / negative
Trigger
Trigger of bursts
AUTO, MANUAL, EXTERN
Synchronization
0 - 360° (16 - 500Hz)
Asynchronous = 0°
a reference signal is connected to the Sync input.
Burst duration td
0.10ms
– 999ms
Burst repetition rate tr
10ms
– 9999ms
Spike frequency f
0.1kHz
– 1000kHz
Range
Step
< 10 kHz
0.1 kHz
10
– 100 kHz
1.0 kHz
100
– 250 kHz
10.0 kHz
> 250 kHz
50.0 kHz
Test duration T
0:01 min - 99:59 min
Output
Direct via 50
coaxial connector
To connect ext. coupling devices
To connect ext. coupling devices
Coupling network
To L, N, PE all combinations
To L, N, PE all combinations
DUT power mains supply
AC 300 V / 16 A / 50/60 Hz
AC 300 V / 16 A / 50/60 Hz
DC 300 V / 160 A
DC 300 V / 16 A
Test routines
Quick Start
Immediate start, all parameters adjustable during a running test
Standard test as per
IEC 61000-4-4 level 1
IEC 61000-4-4 level 2
IEC 61000-4-4 level 3
IEC 61000-4-4 level 4
EN 61000-6-1 Generic
EN 61000-6-2 Generic
IEC 61000- 4-4 Manual operated standard test routine
User test routines
Synchronous burst release
Random burst release
Change level V after T by steps of dV
Frequency sweep in one single burst
Frequency sweep with constant pulses
Frequency sweep, constant burst duration
Change polarity after T
* With Burst pulses as per. IEC 61000-4-4 Ed2 : 2004 the max. output voltage can be limited.