![EMTEST UCS 500 N7 Скачать руководство пользователя страница 76](http://html1.mh-extra.com/html/emtest/ucs-500-n7/ucs-500-n7_manual_2404983076.webp)
EM TEST
UCS 500 Series
Operating Manual
V 5.40
76 / 99
11.8. 50/60Hz Magnetic Field as per IEC 61000-4-8
Magnetic field standard routine IEC 61000-4-8
<- -> O O
5 A/m 01:00 min
START O H/T
F1
F2
F3
F4
F5
F6
F7
With the inc knob the values for the magnetic field strength and test duration can be changed. The actual
parameter is indicated by a blinking circle.
Test setup
Warning
For magnetic field testing the power mains input at PF1 shall be
disconnected.; 230V/16A.
The voltage V is adjusted with a variac as long as the required antenna current is available and the related H field
is generated in the center of the magnetic field antenna.
The variac, type MV 2616 may be controlled automatically via the test generator. Any equivalent variac available
in the lab can be used to control the current manually. Please take care that the variac has a sufficient current
capability.
For more detailed information the operator can require an additional manual especially for magnetic field
testing. This manual is part of the delivery of magnetic field testing accessories.
Test setup with MC 2630 for H-Fields up to 30A/m
Schematics
Test setup
Schematics and Test setup with UCS500Nx, MV 2616, MC 2630 and MS 100N
Option for required magnetic field tests as per IEC 61000-4-8
- External motor variac
MV 2616
- External Magnetic field antenna
MS 100N
- External current transformer
MC 2630
to test 1, 3, 10 and 30A/m levels