Operating Manual
V 5.40
26 / 99
6.3.
Power Fail Generator as per IEC 61000-4-11
EUT supply
Model N5
Model N7
Channel PF1 and PF2
AC voltage/current
max. 300V/16A
max. 300V/16A
Mains frequency
50/60 Hz
DC voltage/current
max. 300V/16A
max. 300V/16A
Inrush current
more than 500A
Protection
Electronic fuse for continuous overcurrent / inrush currents
Electronic control of overheating
PF1 and PF2 are safe against short circuit
max. current channel PF1
nominal current continuous
max. current channel PF2
nominal current with intermittently 10s (50% duty cycle)
Trigger
Events trigger
AUTO, MAN, EXTERN
Repetition rate
0.01s
– 9’999s
Accuracy 0.05% + phase sync
Duration of events
0.02ms
– 9’999s
Accuracy 0.05%
Events preselection
1 - 30'000 or endless
1 - 30'000 or endless
Counter
1 - 1000000
1 - 1000000
Synchronization
0 - 360° (16 - 500Hz)
Asynchronous = 0° if a reference signal is connected to the Sync input.
Max Sync input voltage
Max. 300V or same as EUT voltage specs.
Resolution
1°
Measurements
DUT supply
AC/DC voltage in the LCD display, divider ratio 1:100 +/- 10%
BNC output MON U
Measurement of the EUT supply divider ratio 1:100
5%
BNC output MON I
Measurement of the EUT current and the inrush current
10mV/A
5%, max. 1000A
CRO TRIGGER
positive going flank
0-10V Control Output
0-10V DC for external voltage source
Test routines
Quick Start
Immediate start, all parameters adjustable during a running test
Standard test routines
as per IEC 61000-4-11
ac power ports
as per IEC 61000-4-29
dc power ports
as per IEC 61000-6-1
Generic
as per IEC 61000-6-2
Generic
Manual operated standard test routine
User test routines
Voltage variation, external variac control
Change phase angle W after n events by dA
Change events duration td after n events by dtd
Inverse mode
Magnetic field test
test routine according to IEC 61000-4-8
test level 1, 3, 10 and 30A/m with MC 2630 and a variac
test level 100, 300 and 1000A/m with MC26100
Magnetic field tests per IEC 61000-4-8 and -9
The test routines for handling the magnetic field tests are included in the internal UCS 500N firmware. All functions to
control external options as voltage/current sources or magnetic field antennas are included. In addition the following
hardware is required:
Option required for Magnetic Field Test per IEC 61000-4-9
- Magnetic field antenna (square 1mx1m coil MS 100)
- Adapter for connecting the square coil to the surge output.
Option required for Magnetic Field Test 50/60Hz per IEC 61000-4-8
- External variac (MV2616) and magnetic field antenna (square 1mx1m coil MS 100)
- External current transformer (MC2630) to test 1, 3, 10 and 30 A/m levels
- External current transformer (MC26100) to test level 100, 300 and 1000 A/m levels (short term).