EM TEST
UCS 500 Series
Operating Manual
V 5.40
64 / 99
F7: Voltage Variation
Page 4 ( Show parameter)
Power Fail IEC 61000-4-11
Variation
V2% =
70%
A
= 0 dgr
ts
=
20ms
tr
= 10s
ti
= 0.5s
n
= 3
Counter
START / STOP
000002
F1
F2
F3
F4
F5
F6
F7
Attention: This is the Voltage Variation function which is required in the IEC 61000-4-11:2004. The Voltage
Variation function which is conform with the previous IEC 61000-4-11 version is available in the USER
TEST ROUTINES.
Attention: This is a the voltage variation function which is required in the IEC 61000-4-11: (2004-03) and
simulate a voltage dip during a motor start. For this test requires for the ramp function a motorized
variac MV 2616 is mandatory.
The parameters, defined in the standard are fix. The user can modify the bold marked parameters :
- Phase angle
W [ 0 ... 360 dgr / asynchron]
- Repetition time
tr [5s ... 999s]
- Number of Events
n [1 ... 30'000 / endless]
The Voltage Variation function which is conform with the older IEC 61000-4-11:1995 is available in the menu
User test routines
.