4.
Operation
4.1.
Description of the menus
The simulator is operated by an easy menu control system. Seven function
keys are available to select parameters and functions. All functions are
indicated on the display; max. 8 lines and 40 characters.
The selected parameter is blinking and can be changed by turning the knob (inc./dec.).
:
The digit for change can be selected with the cursor (
).
- Selected values are direct indicated on the screen.
- Status on the bottom lines shows the desired status after pressing the function key.
ESC
:
ESC will take you back to the previous level in the menu and set the displayed
values. The latest settings are stored automatically and will be recalled when the
menu is selected again.
EXIT
: The firmware will reset to the main screen.
EM TEST
UCS 500 N5
Ultra Compact Simulator
V 5.00a02
Burst
Surge
Power Fail
SWN: 001234
The serial number and the version number SWN are
used for traceability reasons. These numbers are listed
in the test reports and calibration certificates. These
numbers also are listed within the test reports
generated by the iec.control software.
Start-up display example UCS 500N5. The models type is displayed after startup.
4.2.
Menu structure
Level 0...4
Level 0
Level 1
Level 2
Level 3
Level 4
MAIN MENU
F1 BURST IEC 61000-4-4
F2 SURGE IEC 61000-4-5/9
F3 PFS
IEC 61000-4-11/8/29
F7 SERVICE
Burst IEC 61000-4-4
F1 Quick Start
F2 Standard routines
F3 User test routines
Quick Start
F1 Start
F2 Change
F3 Continue
Start
Start the test routine
Change
Select all parameters
Continue
Continue the test routine
Surge IEC 61000-4-5/9
Standard routines
Preprogrammed test routines as
per standard requirements
Standard routines
F1 : F4 IEC 61000-4-4 Level 1-4
F5 Generic Standard EN 61000-6-1
F6 Generic standard EN 61000-6-2
F7 Manual standard routine
Standard routines F1..F3
F1 Start
F2 Change
F3 Continue
Power Fail IEC 61000-4-8/11/29
User test routines
Preprogrammed test routines for
evaluation and design support
F1 Synchronous to the mains
F2 Random burst release
F3 Change V after T by
U
F4 Frequency sweep I
F5 Frequency sweep II
F6 Frequency sweep III
F7 Change polarity after T
User test routines F1..F3
F1 Start
F2 Change
F3 Continue
Service
F1 Addresses
F3 Setup
F4 Change standard levels
Setup
F1 Change language
F2 LCD backlighting
F3 Interfaces
F4 ESD/keyboard beeper
F5 Running time clock
F6 Set voltage
F7 Magnetic field factors
Change language Gern or English
LCD backlighting On, Off or Auto
Interfaces Select all parameters
ESD/keyboard beeper (on, off)
Running time clock
Display of the TEST ON time
Set voltage (ext. motor variac)
Magnetic field factors
Correction factors for magnetic field
antenna and current transformer
Change standard level
Change standard level
F1 All parameters to standard level
F2 IEC 61000-4
F3 EN 61000-6-1
F4 EN 61000-6-2
Change standard levels
F1 IEC 61000-4-4 Burst
F2 IEC 61000-4-5 Surge
...