Instructions for Use Oxylog 3000 SW 1.n
99
Cleaning, Disinfection and Sterilization
Reprocessing procedure
– Clean breathing valve, flow sensor, angled
adapter and ventilation hoses of the reusable
hose system after use.
– Always exchange the disposable hose system
after use on a patient.
– The disposable hose system must always be
disposed of correctly. Refer to the chapter "Dis-
posal" on page 107.
– Clean the ventilator and medical gas hoses with
a disposable cloth if heavily soiled.
– Replace the disposable hose system in case of
contamination by e.g. vomitus or coughing.
– Clean the reusable hose system in case of con-
tamination by e.g. vomitus or coughing.
Cleaning and disinfecting
To ensure material compatibility, use disinfectants
based on:
– aldehydes
– alcohols
– quaternary ammonia compounds.
Users in the Federal Republic of Germany are
recommended to use only disinfectants on the
current DGHM list (DGHM: German Society for
Hygiene and Microbiology).
The following disinfectants on the DGHM list are
recommended:
– Dismozon pur
– Incidur
– Sekusept Powder
– Trichlorol
The DGHM list (published by: mhp-Verlag, Wiesba-
den) also specifies the active ingredient in each dis-
infectant. Disinfectants based on the active
ingredients aldehydes, alcohols or quaternary
ammonia compounds are recommended for users
in those countries in which the DGHM list is not
available.
WARNING
Always follow hospital/EMS procedures for
handling equipment contaminated with body
fluids.
WARNING
Always follow local regulations governing the
disposal of infectious waste and materials
contaminated with body fluids to prevent the
risk of infection.
CAUTION
Disinfectants based on:
– compounds containing alkylamine
– compounds containing phenol
– compounds releasing halogen
– strong organic acids
– compounds releasing oxygen
may cause damage, which is not always
immediately apparent, to materials, particularly
those used for the breathing valve, flow sensor
and angled connector.
CAUTION
Sterilization of the ventilator itself by any means,
including ethylene oxide (EtO), is not allowed.
WARNING
Always follow accepted hospital/EMS
procedures for disinfecting equipment
contaminated with body fluids (protective
clothing, eyewear, etc.).
Содержание Oxylog 3000
Страница 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...