Assembly
32
Instructions for Use Oxylog 3000 SW 1.n
Connecting the power supply
The Oxylog 3000 is designed to operate on power
supplies with different voltages.
Internal supply
– With rechargeable battery (specified Smart
Battery, refer to the "Technical Data"
on page 111 for additional information).
Additional external power supply
To recharge the battery and to extend the electrical
operation time.
– DC voltage from the on-board power supply via
DC/DC converter or
– with AC/DC power pack
For more information refer to page 118.
To have a fully charged battery on hand, refer to the
section "Getting Started" on page 37 for additional
information.
Internal supply with rechargeable battery
Replacing the battery
1
Loosen the screw (C) on the battery
compartment cover (B) counterclockwise to
release the cover.
2
Remove the battery cover.
3
Remove the battery (A) by pulling the tab.
Checking the charge of the battery
z
Press the button on the rechargeable battery.
The charge status is indicated as a percentage
by LEDs.
WARNING
A fully charged battery must always be
installed for safety reasons, even when
operating from an external power supply!
WARNING
Treatment of batteries:
– Do not throw into fire,
– Do not force open,
Danger of bodily injury.
015
A
B
C
Содержание Oxylog 3000
Страница 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...