Instructions for Use Oxylog 3000 SW 1.n
115
Technical Data
Measured value display
Airway pressure measurement
Range
0 to 99 mbar
Resolution
1 mbar
Accuracy
±2 mbar
Maximum airway pressure
Ppeak
Positive end expiratory pressure
PEEP
Mean airway pressure
Pmean
Plateau pressure
Pplat
Flow measurement
Minute volume MV
Range
0 to 99 L/min, BTPS
Resolution
0.1 L/min
Accuracy
±15% of measured value, or ±1 L/min, whichever is
greater.
Tidal volume VTe
Range
0 to 5000 mL, BTPS
Resolution
1 mL
Accuracy
±15% of measured value, or ±25 mL, whichever is
greater.
Frequency measurement
Range
0 to 99 /min
Resolution
1 /min
Accuracy
±1 /min
Curve display
Airway pressure Paw (t) –10 to 100 mbar
Flow (t)
–120 to 120 L/min
Содержание Oxylog 3000
Страница 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...