Getting Started
40
Instructions for Use Oxylog 3000 SW 1.n
Checking readiness for operation
– Whenever the ventilator has been serviced or
the ventilation hoses changed.
– At the latest every six months.
The following functions are checked with the menu-
based test:
– Gas supply present.
– Hose system / breathing valve connected and
OK.
– Alarm functions OK.
– Ventilation functions OK.
– Monitor functions OK.
Oxylog 3000 interrupts the test if a fault is detected.
The relevant fault is indicated on the screen.
WARNING
The patient may be endangered if the above
device check is not completed.