Instructions for Use Oxylog 3000 SW 1.n
55
Operation
CPAP, CPAP/ASB (CPAP/PS)
C
ontinuous
P
ositive
A
irway
P
ressure
A
ssisted
S
pontaneous
B
reathing
Spontaneous breathing can be assisted with ASB
and NIV.
Set the ventilation pattern with the controls below
the display:
– Maximum airway pressure
Pmax
.
– O
2
concentration
O
2
.
The following is to be set on the display:
– Positive end expiratory pressure
PEEP
Apnoea ventilation
Apnoea back-up ventilation is only applicable when
using the CPAP mode. In the event of an apnoea,
the ventilator will automatically activate
volume-controlled mandatory ventilation - IPPV
(CMV).
When an apnoea occurs, the device
simultaneously issues an alarm signal and
switches to volume controlled ventilation with the
parameters frequency
fApnoea
, tidal volume
VTApnoea,
and the maximum airway pressure
Pmax
when the apnoea time
TApnoea
has been
reached. The ventilation time ratio I:E is set to
1:1.5. The plateau time
Tplat
is 0. The patient can
breathe spontaneously during apnoea ventilation.
The mandatory frequency
fApnoea
remains
constant.
WARNING
Only use CPAP for patients with sufficient
spontaneous breathing.
Otherwise there is a risk of the patient
receiving insufficient ventilation.
125
37171
09237171
Paw
Flow
Apnoea alarm time
TApnoea
Start
apnoea ventilation
with apnoea alarm
t
t
1
f
Pmax
Содержание Oxylog 3000
Страница 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...