Operation
46
Instructions for Use Oxylog 3000 SW 1.n
Starting operation
The actual screen display may differ in appearance
or configuration.
Switch ON
z
Briefly press the
O
key (A).
The Oxylog 3000 performs a self-test.
– The self-test will be completed in approxi-
mately six seconds.
During the self-test, the system briefly displays the
starting page with the software version and a
prompt for the operator to select the configuration
menu, or to activate the device check by pressing
the rotary knob.
The bar graph indicates the progress of the self-
test.
Upon completion of the self-test, the ventilator
automatically begins ventilation with the default
settings.
The opening display with configured settings is
displayed if the central rotary knob is not pressed.
The manufacturer's default settings are:
– Ventilation mode
IPPV (CMV
).
– Ventilation time ratio
I:E
= 1:1.5.
– Positive end expiratory pressure
PEEP
= 5 mbar.
– Plateau time
Tplat %
= 0%.
–
Trigger
= OFF.
The manufacturer's default settings can be
adjusted in Customer Service Mode. Refer to "Set
startup settings" on page 81.
WARNING
Only use a ventilator that has been cleaned
and successfully tested for operation, to pre-
vent a health risk for the patient and user.
01337170
Oxylog 3000
A
B
11937171
10
437171
Содержание Oxylog 3000
Страница 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...