Instructions for Use Oxylog 3000 SW 1.n
71
Monitoring
Special functions
Manual inspiration / Inspiration hold
Inspiration Hold for volume-controlled strokes:
Regardless of the time at which it is started, an
automatic ventilation stroke can be extended for up
to max. 15 seconds (in IPPV, IPPVAssist, SIMV,
SIMV/ASB).
Or
manual inspiration:
A ventilation stroke can be started manually
between two automatic strokes and held for up to
max. 15 seconds.
The pattern of the manually started ventilation
stroke corresponds with the set ventilation mode.
For IPPV (CMV), SIMV:
Volume-controlled ventilation stroke determined by
the settings »
V
T« and »
Tinsp
«, as well as »
I:E
«.
For BIPAP
®
(PCV+) , BIPAP
®
/ASB (PCV+/PS):
The pressure-controlled ventilation stroke is deter-
mined by the settings
P
insp and
T
insp.
For CPAP/ASB (CPAP/PS):
The pressure-controlled ventilation stroke is deter-
mined by the setting
Δ
ASB
.
To activate Manual inspiration or Inspiration
hold
z
Press key
Insp. hold
(A) for as long as
inspiration is required.
Oxylog 3000 will either extend the momentary
automatic ventilation stroke accordingly or start
a new ventilation stroke and hold it for up to
max. 15 seconds.
100 % O
2
(optional)
To apply 100 % O
2
for 3 minutes regardless of the
current set value.
z
Briefly press key
100 % O
2
(B).
Its LED lights up for 3 minutes.
The set value is resumed by the ventilator upon
expiry of these 3 minutes, or when the
100 % O
2
is
pressed again. The LED dims.
When medical air is selected as input source, the
100 % O
2
function will not deliver pure oxygen.
01237170
Oxylog 3000
A
B
Содержание Oxylog 3000
Страница 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...