Instructions for Use Oxylog 3000 SW 1.n
81
Configuration
Set startup settings
Range of the settings:
The default settings for the parameters are
displayed on the screen when the ventilator is
switched ON. The settings can be adjusted.
Advance to the second page:
1
Select the line
Page
, confirm and turn rotary
knob.
To restore the manufacturer's defaults:
2
Select and confirm line
Set factory default
.
Select hose type
The type of ventilation hose (reusable or
disposable hose system) can be configured.
The gas outlet connectors for the flow measuring
hoses must correspond to the hose type system
selected.
Parameter
Range
Trigger
0 (OFF) 3 to 15 L/min
PEEP
0 to 20 mbar
I:E
3:1 to 1:4
Tinsp
0.2 to 10.0 s
Tplat %
0 to 50%
Δ
ASB
0 to 35 mbar
Ramp
SLOW, STANDARD, FAST
Pinsp
3 to 55 mbar
O
2
-Flow
0 to 15 L/min
NIV
ON, OFF
Tapn
0 (OFF) to 60 s
VTapn
50 to 2000 mL
fapn
12 to 60 bpm
MV-high
2.0 to 41 L/min
MV-low
0.5 to 40 L/min
fspn-high
10 to 100 bpm
Loudness 1/4
to
4/4
Brightness
1/4 to 4/4
09737170
scr
een 21
11
537170
Содержание Oxylog 3000
Страница 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...