Problem Solving
90
Instructions for Use Oxylog 3000 SW 1.n
!!
Int. battery in use
During ventilation, when the
external power source has
been disconnected the
internal battery becomes
the main power source.
When starting ventilation
while using the internal
battery this alarm will not be
issued.
Press the
Alarm Reset
key to
confirm the alarm.
11
!!
Key failed
Technical defect.
To continue ventilation with this
device, verfiy the ventilation
settings and continuously monitor
the device functions.
Call DrägerService.
18
!!!
Leakage
(not in NIV)
The measured expiratory
tidal volume VT is approxi-
mately 40% lower than the
inspiratory value.
Repair leaks in hose system and
possibly in the tube.
Use new flow measuring hoses.
9
Faulty flow sensor.
Replace flow sensor.
The ventilator may not func-
tion properly.
Contact your local DrägerService
for additional support.
!!
Loss of data
Technical defect.
Contact your local DrägerService
for additional support.
20
!!
Loudspeaker
inop
Technical defect.
To continue ventilation with this
device, verfiy the ventilation
settings and continuously monitor
the device functions.
Call DrägerService.
21
!!!
MV high
The upper alarm limit for the
minute volume MV has
been exceeded.
Check patient's condition, check
ventilation pattern, adjust alarm
limits if necessary.
8
Faulty flow sensor.
Replace flow sensor.
The ventilator may not func-
tion properly.
Contact your local DrägerService
for additional support.
Alarm
Cause
Remedy
Alarm
Rank
Содержание Oxylog 3000
Страница 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...