Operation
50
Instructions for Use Oxylog 3000 SW 1.n
Cardio-pulmonary Resuscitation (CPR)
During CPR, the airway pressure
Paw
is increased
because of chest compressions.
The Oxylog 3000 will try to limit the airway pressure
Paw
to the set
Pmax
, without ending the inspiration
prematurely.
However, if due to compressions the airway
pressure
Paw
exceeds the set
Pmax
by 5 mbar,
the Oxylog 3000 cycles to the expiration phase.
Therefore in general, if
Pmax
is set to a higher
value, a higher minute volume is possible.
However, this increases the intra-thoracic pressure
and may reduce coronary perfusion.