Instructions for Use Oxylog 3000 SW 1.n
7
For Your Safety and that of Your Patients
General WARNINGS and CAUTIONS
The following WARNINGS and CAUTIONS apply to
general operation of the device. WARNINGS and
CAUTIONS specific to subsystems or particular
features appear with those topics in later sections
of the manual.
Note on EMC/ESD risk for the device
function
General information on electromagnetic
compatibility (EMC) pursuant to international EMC
standard IEC 60601-1-2:
Electromedical devices are subject to special
precautionary measures concerning
electromagnetic compatibility (EMC) and must be
installed and put into operation in accordance with
the EMC information. Refer to section "Technical
Documentation for the Oxylog 3000 according to
EMC standard IEC/EN 60601-1-2" on page 121.
Portable and mobile RF communications equip-
ment can affect medical electrical equipment.
WARNING
Ventilation monitoring is mandatory at all
times! Whenever a patient is connected to the
ventilator, constant attention by qualified
medical staff is required in order to provide
immediate corrective action in case of a
malfunction.
The operator shall not rely on the built-in
monitoring of the ventilator only and must
always assume full responsibility for proper
ventilation and patient safety in all situations.
WARNING
Keep a manual resuscitation bag available!
If a failure is detected in the ventilator and its
life-support functions can no longer be
guaranteed (e.g. in case of a power failure or
interruption in the medical gas supply),
ventilation must be started without delay with
an independent ventilation device
(resuscitation bag) – using PEEP and/or
increased inspired O
2
concentration as
necessary.
WARNING
Always use officially approved gas cylinders
and pressure regulators that comply with all
applicable regulations.
WARNING
To ensure proper ventilation, always consider
the dead space of the total ventilation circuit
when setting ventilation parameters,
especially when applying small tidal volumes.
When using tidal volume settings below
100 mL Dräger Medical recommends constant
monitoring of etCO
2
and SpO
2
.
Содержание Oxylog 3000
Страница 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...