Assembly
30
Instructions for Use Oxylog 3000 SW 1.n
When using a bacterial filter or HME
z
Connect the bacterial filter or HME to the angled
connector.
Hose connections
1
Connect the ventilation hose (A) to the
breathing valve.
2
Connect the flow measuring hoses (B) to the
nozzles on the flow sensor. Note the different
diameters.
3
Connect the flow measuring (A) hoses to the
Oxylog 3000.
4
Connect the ventilation hose (B) to the gas
outlet on the Oxylog 3000.
NOTE
When using a bacterial filter or HME, measured
flows may deviate from the expiratory flows, as
temperature and humidity of the gas are reduced.
01
0
WARNING
HME and bacterial filters increase the
exhalation resistance and dead space volume
of the ventilation system.
01
1
A
B
B
WARNING
Do not use electrically conductive hoses!
Risk of electric shock.
This can endanger the patient.
012
B
A
Содержание Oxylog 3000
Страница 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...