Principles of Operation
130
Instructions for Use Oxylog 3000 SW 1.n
Using BIPAP
®
As with SIMV, the time pattern is set using the basic
setting parameters of frequency Freq. and
inspiration time Tinsp. The lower pressure level is
set with the PEEP parameter, while the upper level
is set with Pinsp.
When switching over from IPPV to BIPAP
®
mode,
note that the inspiration time Tinsp is set, instead of
the ventilation time ratio I:E.
When switching over from SIMV to BIPAP
®
mode –
while retaining the time pattern – only the Pinsp
setting needs to be changed.
The steepness of the increase from the lower
pressure level to the upper pressure level is
controlled by the
Ramp
setting.
During the lower pressure level phase,
spontaneous breathing can be assisted by ASB.
The steepness of the pressure increase to ASB
pressure
Δ
PASB above PEEP is also controlled by
the
Ramp
setting.
The transition from controlled ventilation via the
weaning phase to fully spontaneous breathing is
achieved by a gradual reduction of inspiratory
pressure Pinsp and/or frequency Freq.
03137171
Paw
PEEP
t
Trigger-
window
Spontaneous
breathing
Mandatory
BIPAP stroke
Synchronized
BIPAP stroke
T
insp
Freq.
1
P
insp
ramp
Содержание Oxylog 3000
Страница 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...