Problem Solving
94
Instructions for Use Oxylog 3000 SW 1.n
Error messages during the device check
Message
Cause
Explanation/Remedy
No communication control- /
charge-board
Device defective
Contact your local DrägerService
for additional support.
System leakage
Leak in ventilation hose system
and/or test lung.
Check hoses, breathing valve,
flow sensor and test lung for leaks
and replace if necessary.
Internal leak in system
Contact your local DrägerService
for additional support.
No test lung
Test lung not connected or major
leakage
Connect test lung.
Check hoses, breathing valve,
flow sensor and test lung for leaks
and replace if necessary.
Breathing valve inop
Breathing valve has malfunc-
tioned.
Check correct condition of breath-
ing valve including diaphragm and
rubber disc; fit a new breathing
valve if necessary or use a new
disposable hose set.
Pressure measurement inop
The ventilation hose system has
not been connected correctly.
Connect ventilation system cor-
rectly.
Pressure measurement is not pos-
sible.
Contact your local DrägerService
for additional support.
PEEP-valve inop
Internal leak in system
Check hoses, breathing valve,
flow sensor and test lung for leaks
and replace if necessary.
Device defective
Contact your local DrägerService
for additional support.
Patient flow measurement
inop
Flow measurement implausible
Replace flow sensor.
Contact your local DrägerService
for additional support.
Содержание Oxylog 3000
Страница 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...