Problem Solving
92
Instructions for Use Oxylog 3000 SW 1.n
!
Self test OK
The device has been
switched on and the self-
test completed successfully.
The message can be confirmed or
it will be cancelled automatically
with the next message.
28
!
Settings not con-
firmed
Parameters have been
changed on the screen but
not confirmed.
Press the rotary knob to confirm
the parameter changes.
27
!!
Supply pressure
low
Supply pressure <270 kPa.
Ensure that supply pressure
exceeds 270 kPa.
11
Alarm
Cause
Remedy
Alarm
Rank