Instructions for Use Oxylog 3000 SW 1.n
89
Problem Solving
!
Check settings
O
2
The set O
2
concentration
cannot be achieved with the
set flow.
Adjust inspiratory flow or O
2
con-
centration (in accordance with
measured value).
27
!!
Check settings
time
The expiration time result-
ing from the settings for
Freq.
and
I:E
or
Tinsp
is not
possible.
Change
Freq.
or
I:E
or
Tinsp
.
13
!!!
Confirm device
OFF with rotary
knob
Key
O
has been pressed
for 3 seconds.
To switch OFF: confirm.
To continue ventilation, press key
O
again.
---
!!!
Device failure
Technical defect.
Contact your local DrägerService
for additional support.
1
!!
Flow measure-
ment inop
Measurement hoses for
flow measurement hoses
kinked, disconnected or
leaking.
Ensure flow measurement hoses
are connected correctly.
22
Flow sensor defective.
Replace flow sensor.
Technical defect.
Disconnect the patient from the
device and continue ventilation
without delay using another
ventilator.
Call DrägerService.
!!
Gas delivery fail-
ure
Technical defect.
Contact your local DrägerService
for additional support.
19
!!
High frequency
Patient breathes at a high
spontaneous rate.
Check patient's condition, check
ventilation pattern, correct alarm
limit
fspn
if necessary.
14
!!
Int. battery
charging inop
Technical defect.
Contact your local DrägerService
for additional support.
16
!!!
Int. battery dis-
charged
The operating time for oper-
ation with the internal bat-
tery has expired and an
external DC supply has not
been connected.
The ventilator must immediately
be reconnected to a mains supply,
an on-board DC supply or a fully
charged battery.
2
Alarm
Cause
Remedy
Alarm
Rank
Содержание Oxylog 3000
Страница 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...