
Technical Data
114
Instructions for Use Oxylog 3000 SW 1.n
Performance data
Control principle
Time-cycled, volume-constant, pressure-controlled.
Maximum inspiratory flow
100 L/min
1)
Device compliance
with 1.5 m ventilation
hose
< 1 mL/mbar
with 3 m ventilation hose < 2 mL/mbar
Inspiration resistance
< 6 mbar at 60 L/min
< 4 mbar at 30 L/min
< 2 mbar at 5 L/min
Expiration resistance
< 6 mbar at 60 L/min
< 4 mbar at 30 L/min
< 2 mbar at 5 L/min
Dead space including flow sensor
approximately 35 mL (reusable hose system).
approximately 33 mL (disposable hose system).
Supplementary functions
Demand valve
Opens the breathing system upon failure of the gas
supply, permits spontaneous breathing with ambient
air.
Relief valve
Opens the breathing system at approximately
80 mbar.
Patient connection
22 mm ISO conical connector.
1) At service pressures >350 kPa.
The maximum inspiratory flow is reduced to 80 L/min at service pressures <350 kPa and to 39 L/min at service
pressures <280 kPa.
Содержание Oxylog 3000
Страница 26: ...26 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 44: ...44 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 74: ...74 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 86: ...86 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 102: ...102 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 106: ...106 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 110: ...110 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 134: ...134 Instructions for Use Oxylog 3000 SW 1 n This page intentionally left blank ...
Страница 137: ...Instructions for Use Oxylog 3000 SW 1 n 137 This page intentionally left blank ...