Inoperative eluent proportioning valve (gradient pump only)
The proportioning valve assembly should be replaced. Contact Dionex for
assistance.
•
Inoperative check valves
Replace the check valve cartridges (see
•
Insufficient mixing (gradient pump only)
The static mixer may be dirty or contaminated. Flush with IPA (isopropyl
alcohol). If this does not eliminate the problem, install a new mixer (GM-3,
P/N 042126; GM-4, P/N 049135).
•
Eluent is contaminated
1. Clean all eluent reservoirs thoroughly (inside and out) with ASTM Type I
(18 megohm-cm) filtered and deionized water and dry with clean,
particulate-free air. If a reservoir still appears dirty, or if there is a slimy
film on the interior, follow the cleaning instructions in
.
2. Flush the system with ASTM Type I (18 megohm-cm) filtered and
deionized water.
3. Replace all end-line filters (P/N 045987). See
for instructions.
4. Prepare new stock solution.
5. Prepare fresh eluent. To ensure eluent purity, prepare all eluents with
electrochemical-grade chemicals, and ASTM Type I (18 megohm-cm)
filtered and deionized water.
•
Problem unrelated to the pump
1. The eluent concentration may be wrong, or the eluent may have been
prepared with impure chemicals/eluents or water. To ensure eluent purity,
prepare all eluents with spectro-grade eluents, reagent-grade chemicals,
and ASTM Type I (or better) filtered and deionized water.
2. The column may be the source of the problem. Refer to the column
manual for troubleshooting assistance. Column manuals are provided on
the Dionex Reference Library CD-ROM (P/N 053891).
3. The injection valve may be the source of the problem. Rebuild the
injection valve (see
Содержание ICS-3000
Страница 1: ...ICS 3000 Ion Chromatography System Operator s Manual Document No 065031 Revision 04 January 2008...
Страница 32: ...ICS 3000 Ion Chromatography System 16 Doc 065031 04 1 08...
Страница 93: ...2 DC Description Doc 065031 04 1 08 77 Figure 2 42 DC Control Panel...
Страница 99: ...2 DC Description Doc 065031 04 1 08 83 Figure 2 46 Conductivity Detector Control Panel...
Страница 106: ...ICS 3000 Ion Chromatography System 90 Doc 065031 04 1 08...
Страница 138: ...ICS 3000 Ion Chromatography System 122 Doc 065031 04 1 08...
Страница 166: ...ICS 3000 Ion Chromatography System 150 Doc 065031 04 1 08...
Страница 174: ...ICS 3000 Ion Chromatography System 158 Doc 065031 04 1 08...
Страница 180: ...ICS 3000 Ion Chromatography System 164 Doc 065031 04 1 08...
Страница 218: ...ICS 3000 Ion Chromatography System 202 Doc 065031 04 1 08...
Страница 228: ...ICS 3000 Ion Chromatography System 212 Doc 065031 04 1 08...
Страница 272: ...ICS 3000 Ion Chromatography System 256 Doc 065031 04 1 08...
Страница 296: ...ICS 3000 Ion Chromatography System 280 Doc 065031 04 1 08...
Страница 354: ...ICS 3000 Ion Chromatography System 338 Doc 065031 04 1 08...
Страница 362: ...ICS 3000 Ion Chromatography System 346 Doc 065031 04 1 08...
Страница 368: ...ICS 3000 Ion Chromatography System 352 Doc 065031 04 1 08...
Страница 372: ...ICS 3000 Ion Chromatography System 356 Doc 065031 04 1 08...
Страница 374: ...ICS 3000 Ion Chromatography System 358 Doc 065031 04 1 08...