8 • Troubleshooting
Doc. 065031-04 1/08
181
2. Air bubbles may be trapped inside the cell. While wearing gloves and eye
protection, generate a slight temporary backpressure by putting your
finger over the end of the cell outlet tubing for 2 to 3 seconds. Repeat 2 or
3 times. If the baseline does not improve, check the other causes of
baseline instability described in this section.
NOTE To prevent air from becoming trapped in the cell in
the future, increase the backpressure on the cell by
connecting backpressure tubing to the cell outlet.
The backpressure limit for the ED
cell
is 690 kPa
(100 psi). Do not exceed this limit.
3. The water used to prepare the eluent contains trace contaminants. Remake
the eluent, using ASTM Type I (18 megohm-cm) filtered and deionized
water containing no contaminants.
8.3
Poor Retention Time Reproducibility
•
Liquid leaks
1. Check for leaks from the piston seals. Replace the piston seal on any head
with a leak (see
2. Check for leaks throughout the remainder of the system, including the
check valves, injection valve, and columns. Tighten or replace fittings as
needed.
•
DP/SP not primed
•
Liquid lines incompletely flushed after an eluent change
Attach a 10 cc syringe (P/N 054578) to the waste port on the priming valve.
Open the priming valve (turn the knob one-half turn counterclockwise). Draw
at least 20 mL of the new eluent through the liquid lines before operation.
Do not block the end of the cell outlet tubing for longer than the
recommended 2 to 3 seconds. Doing so creates high backpressure,
which can break the reference electrode glass membrane.
Содержание ICS-3000
Страница 1: ...ICS 3000 Ion Chromatography System Operator s Manual Document No 065031 Revision 04 January 2008...
Страница 32: ...ICS 3000 Ion Chromatography System 16 Doc 065031 04 1 08...
Страница 93: ...2 DC Description Doc 065031 04 1 08 77 Figure 2 42 DC Control Panel...
Страница 99: ...2 DC Description Doc 065031 04 1 08 83 Figure 2 46 Conductivity Detector Control Panel...
Страница 106: ...ICS 3000 Ion Chromatography System 90 Doc 065031 04 1 08...
Страница 138: ...ICS 3000 Ion Chromatography System 122 Doc 065031 04 1 08...
Страница 166: ...ICS 3000 Ion Chromatography System 150 Doc 065031 04 1 08...
Страница 174: ...ICS 3000 Ion Chromatography System 158 Doc 065031 04 1 08...
Страница 180: ...ICS 3000 Ion Chromatography System 164 Doc 065031 04 1 08...
Страница 218: ...ICS 3000 Ion Chromatography System 202 Doc 065031 04 1 08...
Страница 228: ...ICS 3000 Ion Chromatography System 212 Doc 065031 04 1 08...
Страница 272: ...ICS 3000 Ion Chromatography System 256 Doc 065031 04 1 08...
Страница 296: ...ICS 3000 Ion Chromatography System 280 Doc 065031 04 1 08...
Страница 354: ...ICS 3000 Ion Chromatography System 338 Doc 065031 04 1 08...
Страница 362: ...ICS 3000 Ion Chromatography System 346 Doc 065031 04 1 08...
Страница 368: ...ICS 3000 Ion Chromatography System 352 Doc 065031 04 1 08...
Страница 372: ...ICS 3000 Ion Chromatography System 356 Doc 065031 04 1 08...
Страница 374: ...ICS 3000 Ion Chromatography System 358 Doc 065031 04 1 08...