ICS-3000 Ion Chromatography System
270
Doc. 065031-04 1/08
e. The seal wash seal is now partially installed. To complete the seal
installation, place the retainer in the seal wash body and use the
large flat-blade screwdriver to tighten the retainer.
f. Place the new O-ring (P/N 040695) on the seal wash body.
NOTE When replacing a piston seal, always replace
the O-ring, also. This will prevent leaks.
2. Remove the 10-32 fitting plugs from the pump head.
3. Insert the piston through the seal wash assembly, and then through the
new piston seal (P/N 064946).
4. Make sure the piston seal is centered.
5. Rinse the piston seal with isopropyl alcohol (IPA) or dip it into a
container of IPA. (The seal is easier to install when it is moist.)
6. Place the front of the pump head, flat side down, on a clean work
surface.
Make sure the open side of the piston seal faces away from the
retainer for the seal wash seal.
7. Place the components on the pump head and
gently
press the housing
until the piston seal snaps into place.
9.5.6
Reinstalling the Piston and Pump Head
1. Slide the piston
partway
into the pump head; approximately 6 mm
(1/4 in) of the sapphire part of the piston should extend from the head.
2. Place the pump head back on the pump.
3. Reinstall the screws in the pump head. Using the 3.0 mm hex key
(P/N 062338), tighten the screws just until they come into contact
If the piston seal is not centered, applying pressure to it in
will
damage the seal and make it unusable.
Do not use a sharp tool (such as tweezers) to install the piston seal.
This will scratch the seal and the inside of the pump housing; these
scratches will prevent a proper seal and cause leakage.
Содержание ICS-3000
Страница 1: ...ICS 3000 Ion Chromatography System Operator s Manual Document No 065031 Revision 04 January 2008...
Страница 32: ...ICS 3000 Ion Chromatography System 16 Doc 065031 04 1 08...
Страница 93: ...2 DC Description Doc 065031 04 1 08 77 Figure 2 42 DC Control Panel...
Страница 99: ...2 DC Description Doc 065031 04 1 08 83 Figure 2 46 Conductivity Detector Control Panel...
Страница 106: ...ICS 3000 Ion Chromatography System 90 Doc 065031 04 1 08...
Страница 138: ...ICS 3000 Ion Chromatography System 122 Doc 065031 04 1 08...
Страница 166: ...ICS 3000 Ion Chromatography System 150 Doc 065031 04 1 08...
Страница 174: ...ICS 3000 Ion Chromatography System 158 Doc 065031 04 1 08...
Страница 180: ...ICS 3000 Ion Chromatography System 164 Doc 065031 04 1 08...
Страница 218: ...ICS 3000 Ion Chromatography System 202 Doc 065031 04 1 08...
Страница 228: ...ICS 3000 Ion Chromatography System 212 Doc 065031 04 1 08...
Страница 272: ...ICS 3000 Ion Chromatography System 256 Doc 065031 04 1 08...
Страница 296: ...ICS 3000 Ion Chromatography System 280 Doc 065031 04 1 08...
Страница 354: ...ICS 3000 Ion Chromatography System 338 Doc 065031 04 1 08...
Страница 362: ...ICS 3000 Ion Chromatography System 346 Doc 065031 04 1 08...
Страница 368: ...ICS 3000 Ion Chromatography System 352 Doc 065031 04 1 08...
Страница 372: ...ICS 3000 Ion Chromatography System 356 Doc 065031 04 1 08...
Страница 374: ...ICS 3000 Ion Chromatography System 358 Doc 065031 04 1 08...