ICS-3000 Ion Chromatography System
228
Doc. 065031-04 1/08
When a TC is shared by two timebases, the same values must be selected for
certain parameters. If the Chromeleon Ready Check detects a discrepancy,
this error occurs and batch processing cannot be started.
To troubleshoot:
Find and eliminate any discrepancies in the parameter settings. For example,
all samples must turn on
TempCtrl
or all samples must turn off
TempCtrl
.
Selecting the
Disconnect
or
Standby
command during data acquisition will
stop data transmission and cause data acquisition to stall. If you attempt to
issue one of these commands during data acquisition, this error occurs.
To troubleshoot:
Wait until data acquisition is completed before attempting to issue either of
these commands.
All samples scheduled for run on timebases x need to specify
the same value for property (nominal) Temperature.
-or-
All samples scheduled for run on timebases x need to specify
the same value for property TempCtrl.
-or-
All samples scheduled for run on timebases x need to specify
the same value for property y.
Can't execute “Disconnect” command during data
acquisition.
-or-
Can't execute “Standby” command during data acquisition.
Содержание ICS-3000
Страница 1: ...ICS 3000 Ion Chromatography System Operator s Manual Document No 065031 Revision 04 January 2008...
Страница 32: ...ICS 3000 Ion Chromatography System 16 Doc 065031 04 1 08...
Страница 93: ...2 DC Description Doc 065031 04 1 08 77 Figure 2 42 DC Control Panel...
Страница 99: ...2 DC Description Doc 065031 04 1 08 83 Figure 2 46 Conductivity Detector Control Panel...
Страница 106: ...ICS 3000 Ion Chromatography System 90 Doc 065031 04 1 08...
Страница 138: ...ICS 3000 Ion Chromatography System 122 Doc 065031 04 1 08...
Страница 166: ...ICS 3000 Ion Chromatography System 150 Doc 065031 04 1 08...
Страница 174: ...ICS 3000 Ion Chromatography System 158 Doc 065031 04 1 08...
Страница 180: ...ICS 3000 Ion Chromatography System 164 Doc 065031 04 1 08...
Страница 218: ...ICS 3000 Ion Chromatography System 202 Doc 065031 04 1 08...
Страница 228: ...ICS 3000 Ion Chromatography System 212 Doc 065031 04 1 08...
Страница 272: ...ICS 3000 Ion Chromatography System 256 Doc 065031 04 1 08...
Страница 296: ...ICS 3000 Ion Chromatography System 280 Doc 065031 04 1 08...
Страница 354: ...ICS 3000 Ion Chromatography System 338 Doc 065031 04 1 08...
Страница 362: ...ICS 3000 Ion Chromatography System 346 Doc 065031 04 1 08...
Страница 368: ...ICS 3000 Ion Chromatography System 352 Doc 065031 04 1 08...
Страница 372: ...ICS 3000 Ion Chromatography System 356 Doc 065031 04 1 08...
Страница 374: ...ICS 3000 Ion Chromatography System 358 Doc 065031 04 1 08...