8.28.2 Cannot Set Amperometry Cell pH Readout to 7.0
The pH reading is displayed on the Chromeleon or Chromeleon Xpress
EC Detector
Control panel.
•
Inaccurate calibration buffer
Use a pH meter to check the pH of the buffer.
•
Dry reference electrode
1. Soak the reference electrode in a solution containing 1 M KCl and
1 M HCl long
enough to restore the electrode potential to
<30 mV when compared to an unexposed Ag/AgCl reference
electrode
. To test the reference electrode potential, see
Digital Voltmeter to Determine Reference Potential Shift” on
page 226
2. Replace the reference electrode (see
To prevent a reference electrode from drying out, make sure that
eluent is being pumped continuously through the cell. If the cell will
not be used for a short time (less than 2 days), disconnect the tubing
from the inlet and outlet fittings and install fitting plugs. For longer
shutdowns, remove the electrode from the cell and store it in its
storage bottle filled with saturated KCl solution. See
detailed instructions.
8.28.3 Shift in Amperometry Cell pH Readout
The pH readout is considered to have shifted if it is 0.5 pH units or more
different from the value observed when the reference electrode was new.
•
Faulty reference electrode
1. Check the electrode by following the instructions in
2. Regenerate the electrode by soaking in a solution containing 1 M KCl
and 1 M HCl long
enough to restore the electrode potential to
<30 mV when compared to an unexposed Ag/AgCl reference
electrode
. To test the reference electrode potential, see
Digital Voltmeter to Determine Reference Potential Shift” on
page 226
Содержание ICS-3000
Страница 1: ...ICS 3000 Ion Chromatography System Operator s Manual Document No 065031 Revision 04 January 2008...
Страница 32: ...ICS 3000 Ion Chromatography System 16 Doc 065031 04 1 08...
Страница 93: ...2 DC Description Doc 065031 04 1 08 77 Figure 2 42 DC Control Panel...
Страница 99: ...2 DC Description Doc 065031 04 1 08 83 Figure 2 46 Conductivity Detector Control Panel...
Страница 106: ...ICS 3000 Ion Chromatography System 90 Doc 065031 04 1 08...
Страница 138: ...ICS 3000 Ion Chromatography System 122 Doc 065031 04 1 08...
Страница 166: ...ICS 3000 Ion Chromatography System 150 Doc 065031 04 1 08...
Страница 174: ...ICS 3000 Ion Chromatography System 158 Doc 065031 04 1 08...
Страница 180: ...ICS 3000 Ion Chromatography System 164 Doc 065031 04 1 08...
Страница 218: ...ICS 3000 Ion Chromatography System 202 Doc 065031 04 1 08...
Страница 228: ...ICS 3000 Ion Chromatography System 212 Doc 065031 04 1 08...
Страница 272: ...ICS 3000 Ion Chromatography System 256 Doc 065031 04 1 08...
Страница 296: ...ICS 3000 Ion Chromatography System 280 Doc 065031 04 1 08...
Страница 354: ...ICS 3000 Ion Chromatography System 338 Doc 065031 04 1 08...
Страница 362: ...ICS 3000 Ion Chromatography System 346 Doc 065031 04 1 08...
Страница 368: ...ICS 3000 Ion Chromatography System 352 Doc 065031 04 1 08...
Страница 372: ...ICS 3000 Ion Chromatography System 356 Doc 065031 04 1 08...
Страница 374: ...ICS 3000 Ion Chromatography System 358 Doc 065031 04 1 08...