9.12.3 Disposing of the Old EluGen Cartridge
If the EluGen cartridge is expended, follow these steps to dispose of it:
1. Hold the cartridge upright. Unscrew the eluent generation chamber
from the electrolyte reservoir and pour the remaining electrolyte
solution into an appropriate hazardous waste container.
NOTE Refer to the Material Safety Data Sheet (MSDS)
shipped with the EluGen cartridge for the
chemical description.
2. Rinse the electrolyte reservoir and membranes with ASTM Type I
(18 megohm-cm) filtered and deionized water
three times
. Rinsing
should render the reservoir and membranes nonhazardous; however,
check your local, state, and federal regulatory agency regulations for
proper disposal.
If the EluGen cartridge is not expended, it can be stored for up to two
years. Follow these steps to prepare the cartridge for storage:
1. Plug all fittings. Cap the vent port (this will prevent the evaporation
of water in the electrolyte reservoir).
2. Store the cartridge in an upright position (with the electrolyte
reservoir at top) at 4 to 40°C (39 to 104°F) until its next use. The
original shipping container is ideal for storage. Before resuming
operation, condition the cartridge. For instructions, see
the New EluGen Cartridge” on page 288
.
9.12.4 Installing the New EluGen Cartridge
1. Remove the new EluGen cartridge from the shipping container.
NOTE Save the shipping container; it may be required
for storage or disposal of the EluGen cartridge.
2. Verify that a 10-32 plug is installed in the EluGen cartridge vent
opening. (This will prevent leakage from the vent opening when you
turn over the cartridge.)
3. Turn the cartridge upside down and place it on the EG drip tray in the
service position (see
Содержание ICS-3000
Страница 1: ...ICS 3000 Ion Chromatography System Operator s Manual Document No 065031 Revision 04 January 2008...
Страница 32: ...ICS 3000 Ion Chromatography System 16 Doc 065031 04 1 08...
Страница 93: ...2 DC Description Doc 065031 04 1 08 77 Figure 2 42 DC Control Panel...
Страница 99: ...2 DC Description Doc 065031 04 1 08 83 Figure 2 46 Conductivity Detector Control Panel...
Страница 106: ...ICS 3000 Ion Chromatography System 90 Doc 065031 04 1 08...
Страница 138: ...ICS 3000 Ion Chromatography System 122 Doc 065031 04 1 08...
Страница 166: ...ICS 3000 Ion Chromatography System 150 Doc 065031 04 1 08...
Страница 174: ...ICS 3000 Ion Chromatography System 158 Doc 065031 04 1 08...
Страница 180: ...ICS 3000 Ion Chromatography System 164 Doc 065031 04 1 08...
Страница 218: ...ICS 3000 Ion Chromatography System 202 Doc 065031 04 1 08...
Страница 228: ...ICS 3000 Ion Chromatography System 212 Doc 065031 04 1 08...
Страница 272: ...ICS 3000 Ion Chromatography System 256 Doc 065031 04 1 08...
Страница 296: ...ICS 3000 Ion Chromatography System 280 Doc 065031 04 1 08...
Страница 354: ...ICS 3000 Ion Chromatography System 338 Doc 065031 04 1 08...
Страница 362: ...ICS 3000 Ion Chromatography System 346 Doc 065031 04 1 08...
Страница 368: ...ICS 3000 Ion Chromatography System 352 Doc 065031 04 1 08...
Страница 372: ...ICS 3000 Ion Chromatography System 356 Doc 065031 04 1 08...
Страница 374: ...ICS 3000 Ion Chromatography System 358 Doc 065031 04 1 08...