TC Description
Doc. 065031-04 1/08
99
•
In the Inject position, sample is swept to the column for analysis.
Eluent flows from the pump, through the sample loop, and on to the
column, carrying the contents of the sample loop with it.
describes how to inject samples.
The 10-port injection valve has two operating positions: 10_1 and 1_2.
Liquid flows through the path between either ports 10 and 1 or ports 1 and
2, depending on the valve position.
shows example flow schematics for the 10-port valve when it
is connected for an application that uses a concentrator column.
•
In the 10_1 position, sample flows from the syringe or autosampler
line, through the valve, and into the sample loop. Excess sample
flows out to waste. Eluent flows from the eluent pump, through the
valve, through the concentrator column, and to the separator column.
If sample was previously loaded onto the concentrator column, it is
swept to the separator column for analysis.
•
In the 1_2 position, carrier liquid flows through the sample loop and
on to the concentrator column, carrying the contents of the sample
Figure 2-56. TC Injection Valve Flow Schematics (10-Port Valve)
Example Connections: Large Loop to Concentrator
1
3
5
7
9
10_1 POSITION
(Loop Loading)
To Waste
Eluent
Pump In
To Column
Sample In
To Waste
Concentrator
Large
Loop
Carrier Pump In
1_2 POSITION
(Concentrator Loading)
To Waste
Eluent
Pump In
To Column
Sample In
To Waste
Concentrator
Carrier Pump In
= Sample
= Carrier
= Eluent
Large
Loop
2
4
6
8
10
1
3
5
7
9
2
4
6
8
10
Содержание ICS-3000
Страница 1: ...ICS 3000 Ion Chromatography System Operator s Manual Document No 065031 Revision 04 January 2008...
Страница 32: ...ICS 3000 Ion Chromatography System 16 Doc 065031 04 1 08...
Страница 93: ...2 DC Description Doc 065031 04 1 08 77 Figure 2 42 DC Control Panel...
Страница 99: ...2 DC Description Doc 065031 04 1 08 83 Figure 2 46 Conductivity Detector Control Panel...
Страница 106: ...ICS 3000 Ion Chromatography System 90 Doc 065031 04 1 08...
Страница 138: ...ICS 3000 Ion Chromatography System 122 Doc 065031 04 1 08...
Страница 166: ...ICS 3000 Ion Chromatography System 150 Doc 065031 04 1 08...
Страница 174: ...ICS 3000 Ion Chromatography System 158 Doc 065031 04 1 08...
Страница 180: ...ICS 3000 Ion Chromatography System 164 Doc 065031 04 1 08...
Страница 218: ...ICS 3000 Ion Chromatography System 202 Doc 065031 04 1 08...
Страница 228: ...ICS 3000 Ion Chromatography System 212 Doc 065031 04 1 08...
Страница 272: ...ICS 3000 Ion Chromatography System 256 Doc 065031 04 1 08...
Страница 296: ...ICS 3000 Ion Chromatography System 280 Doc 065031 04 1 08...
Страница 354: ...ICS 3000 Ion Chromatography System 338 Doc 065031 04 1 08...
Страница 362: ...ICS 3000 Ion Chromatography System 346 Doc 065031 04 1 08...
Страница 368: ...ICS 3000 Ion Chromatography System 352 Doc 065031 04 1 08...
Страница 372: ...ICS 3000 Ion Chromatography System 356 Doc 065031 04 1 08...
Страница 374: ...ICS 3000 Ion Chromatography System 358 Doc 065031 04 1 08...