ICS-3000 Ion Chromatography System
180
Doc. 065031-04 1/08
Noisy Baseline: ED Only
•
(DC Amperometry and Integrated Amperometry modes) Air bubbles
trapped inside cell
While wearing gloves and eye protection and with the pump running and all
plumbing connected, generate a slight temporary backpressure by putting
your finger over the end of the cell outlet tubing for 2 to 3 seconds. Repeat 2
or 3 times. If the baseline does not improve, check the other causes of baseline
instability described in this section.
NOTE To prevent air from becoming trapped in the cell in the
future, increase the backpressure on the cell by
connecting backpressure tubing to the cell outlet. The
backpressure limit for the ED
cell
is 690 kPa (100 psi).
Do not exceed this limit.
•
(DC Amperometry and Integrated Amperometry modes) Frequent,
random spikes in the baseline
The reference electrode diaphragm is plugged. First, try regenerating the
reference electrode frit by soaking the electrode in a solution of 1 M KCl plus
1 M HCl. If this does not eliminate the spiking, replace the electrode
(P/N 061879).
•
(DC Amperometry and Integrated Amperometry modes) Regular
baseline oscillation on high-sensitivity ranges
Reconnect the short length of titanium tubing to the cell inlet (see
•
(DC Amperometry and Integrated Amperometry modes) Dirty or pitted
conventional (nondisposable) working electrode
Polish the working electrode (see
). If you are using a
disposable electrode, replace it.
•
(Integrated Amperometry mode) Regular baseline oscillations
1. Check the pump pressure plot to verify that the pump is working properly.
Do not block the end of the cell outlet tubing for longer than the
recommended 2 to 3 seconds. Doing so creates high backpressure,
which can break the reference electrode glass membrane.
Содержание ICS-3000
Страница 1: ...ICS 3000 Ion Chromatography System Operator s Manual Document No 065031 Revision 04 January 2008...
Страница 32: ...ICS 3000 Ion Chromatography System 16 Doc 065031 04 1 08...
Страница 93: ...2 DC Description Doc 065031 04 1 08 77 Figure 2 42 DC Control Panel...
Страница 99: ...2 DC Description Doc 065031 04 1 08 83 Figure 2 46 Conductivity Detector Control Panel...
Страница 106: ...ICS 3000 Ion Chromatography System 90 Doc 065031 04 1 08...
Страница 138: ...ICS 3000 Ion Chromatography System 122 Doc 065031 04 1 08...
Страница 166: ...ICS 3000 Ion Chromatography System 150 Doc 065031 04 1 08...
Страница 174: ...ICS 3000 Ion Chromatography System 158 Doc 065031 04 1 08...
Страница 180: ...ICS 3000 Ion Chromatography System 164 Doc 065031 04 1 08...
Страница 218: ...ICS 3000 Ion Chromatography System 202 Doc 065031 04 1 08...
Страница 228: ...ICS 3000 Ion Chromatography System 212 Doc 065031 04 1 08...
Страница 272: ...ICS 3000 Ion Chromatography System 256 Doc 065031 04 1 08...
Страница 296: ...ICS 3000 Ion Chromatography System 280 Doc 065031 04 1 08...
Страница 354: ...ICS 3000 Ion Chromatography System 338 Doc 065031 04 1 08...
Страница 362: ...ICS 3000 Ion Chromatography System 346 Doc 065031 04 1 08...
Страница 368: ...ICS 3000 Ion Chromatography System 352 Doc 065031 04 1 08...
Страница 372: ...ICS 3000 Ion Chromatography System 356 Doc 065031 04 1 08...
Страница 374: ...ICS 3000 Ion Chromatography System 358 Doc 065031 04 1 08...