If the leaking check valve is loose, follow these steps to tighten it:
1. Turn off the pump.
2. Loosen the check valve enough to allow it to turn freely.
3. Tighten the check valve fingertight, and then tighten it one-quarter turn
with a 1/2-inch wrench.
4. If the check valve continues to leak, tighten it an additional one-quarter
turn with the 1/2-inch wrench.
5. If the check valve continues to leak, it is defective. Replace both check
•
Eluent proportioning valve leaks (gradient pump only)
Tighten loose fittings. If there are no loose fittings, the proportioning valve
assembly should be replaced. Contact Dionex for assistance.
•
Priming valve knob leaks
If leaks occur when the priming valve knob is open, close the knob
completely
and then open it one-half to three-quarters turn. If this does not stop the
leakage, replace the seal in the priming valve knob (see
If leaks occurs when the priming valve knob is closed, either the pump head
or the knob is damaged and should be replaced. Contact Dionex for
assistance.
•
Excessive system backpressure
If system backpressure is substantially higher than the normal operating
backpressure for the currently configured system (including the column),
tubing may be plugged or overtightened. Follow the procedure below to
isolate segments of the flow path and determine the sources of the high
backpressure.
1. Remove the pump outlet tubing at the injection valve.
2. Press the
POWER
button on the front of the DP/SP to turn on the pump.
3. Set the
Flow
rate to 1.0 mL/min on the Chromeleon or Chromeleon
Xpress Control panel and record the backpressure.
4. One at a time, reconnect each component (tubing, fittings, etc.) of the
flow path. If reconnecting a component causes an abnormal increase in
Содержание ICS-3000
Страница 1: ...ICS 3000 Ion Chromatography System Operator s Manual Document No 065031 Revision 04 January 2008...
Страница 32: ...ICS 3000 Ion Chromatography System 16 Doc 065031 04 1 08...
Страница 93: ...2 DC Description Doc 065031 04 1 08 77 Figure 2 42 DC Control Panel...
Страница 99: ...2 DC Description Doc 065031 04 1 08 83 Figure 2 46 Conductivity Detector Control Panel...
Страница 106: ...ICS 3000 Ion Chromatography System 90 Doc 065031 04 1 08...
Страница 138: ...ICS 3000 Ion Chromatography System 122 Doc 065031 04 1 08...
Страница 166: ...ICS 3000 Ion Chromatography System 150 Doc 065031 04 1 08...
Страница 174: ...ICS 3000 Ion Chromatography System 158 Doc 065031 04 1 08...
Страница 180: ...ICS 3000 Ion Chromatography System 164 Doc 065031 04 1 08...
Страница 218: ...ICS 3000 Ion Chromatography System 202 Doc 065031 04 1 08...
Страница 228: ...ICS 3000 Ion Chromatography System 212 Doc 065031 04 1 08...
Страница 272: ...ICS 3000 Ion Chromatography System 256 Doc 065031 04 1 08...
Страница 296: ...ICS 3000 Ion Chromatography System 280 Doc 065031 04 1 08...
Страница 354: ...ICS 3000 Ion Chromatography System 338 Doc 065031 04 1 08...
Страница 362: ...ICS 3000 Ion Chromatography System 346 Doc 065031 04 1 08...
Страница 368: ...ICS 3000 Ion Chromatography System 352 Doc 065031 04 1 08...
Страница 372: ...ICS 3000 Ion Chromatography System 356 Doc 065031 04 1 08...
Страница 374: ...ICS 3000 Ion Chromatography System 358 Doc 065031 04 1 08...