RG
Sensor Interface
On the sloping front panel of the DT80 there are two rows of terminal blocks – digital channels on the left, analog
channels on the right. The green terminal blocks can be quickly unplugged from the DT80 without unscrewing the sensor
cabling. This interface includes:
•
8 digital input/ output/ counter channels (
1D–8D
), 4 of which are SDI-12 compatible (DT82I: 4 channels; DT81/
82E: 4 channels, one of which is SDI-12 compatible)
•
an input to wake the DT80 from low power "sleep" mode (
WK
)
•
4 high speed counter inputs (7 on DT85/ 85L Series 3) (
1C–4C
)
•
2 phase encoder inputs, shared with the counter inputs connected) (
1PE–2PE
) (one phase encoder input on
DT81, none on DT82E, 3 on DT85/ 85L Series 3)
•
a pair of voltage free relay contact outputs (
RELAY A
and
B
)
•
a general purpose switched 12V 150mA power output (
12V
) (DT80/ 81/ 82E and DT85 Series 2/3 only)
•
a general purpose switched 12V/5V 300mA power output (
12V
/
5V
) (DT80/ 82E, DT85 Series 4 only)
•
a general purpose current limited (300mA) power output (
PWR OUT
), which is derived from the external power
input (DT85 only)
•
an isolated switched 5V power output (
5V SW
) (Series 3 only)
•
an isolated programmable 16-bit voltage and current output (
V/I DAC
) (Series 4 only)
•
digital / power ground terminals (
DGND
)
•
an RS232/422/485 compatible serial port (
Tx
,
Rx
,
RTS
and
CTS
) (not present on DT81/ 82E)
•
a number of analog input channels (5 channels for DT80, 1 channel for DT81, 16 channels for DT85)
•
an external excitation input (
EXT *
)
•
isolated analog ground terminals (
AGND
on DT80/81,
EXT#
on DT80 Series 2 and DT85)
Note:
early production DT80 models only had 4 analog inputs
Communications/ Power Interface
On the left side panel you have a variety of connectivity options:
•
10-Base-T Ethernet for connection to a host computer or local area network
•
USB for high speed connection to a host computer (not present on DT82)
•
RS232 for connection to host computer or modem (not present on DT8xM)
•
two alternative DC power connectors – a standard plug-pack socket (DC jack) and a 4-pin terminal block (2-pin
for DT82E/ 80L/ 85L)
For more details, see
Wireless Modem Interface (DT8xM only)
On the right side panel are the integrated modem interface:
•
SIM card slot (subscriber identity module)
•
red status LED
•
coaxial screw connector for optional "receive diversity" antenna (DT8xM3 only)
•
coaxial screw connector for main antenna.
A threaded earth point is also available on both left and right side panels. See
.
WiFi Interface (DT8xW only)
On the right side panel are the integrated WiFi interface:
•
coaxial screw connector for main antenna.
A threaded earth point is also available on both left and right side panels. See
.
Getting Started
Power
) discusses the ways to provide power to the DT80. The simplest option is to plug in the
supplied AC adaptor.
All DT80 models except the DT82E include an internal 6V lead-acid battery which can power the logger if the main
external supply is interrupted.
Important:
The DT80 is shipped with its main internal battery disconnected. We recommend the battery is connected as soon as
practical so that it can charge from the mains adaptor or other external power source. This is achieved by simply plugging the green