Using the Monitor Ports
7-8
Operator's Manual
Figure 7-2.
Digital Output Connector Pin-Out
7.4.1
Using a Philips VueLink™* Open Interface (VOI) Module
Note:
The INVOS™ system will also interface with Hewlett Packard and Agilent multi-parameter systems that
accept the VOI B module specified in this section.
WARNING:
The VueLink™* monitor will not generate an alarm or error message if remote communication
between the VueLink™* monitor and the INVOS™ system has been broken. During this period of
no remote communication, the INVOS™ system will continue to monitor, alarm, and display status
messages. The VueLink™* monitor operator should not rely upon the distributed alarm system for
generating alarms.
In Run function, the digital output port may be set to output real-time data to communicate with
a Philips VueLink™* Open Interface (VOI) module for display on a Philips VueLink™* multi-param-
eter system. The data sent to the multi-parameter includes the rSO
2
values, alarms, and status
messages (see section
Run Function (Patient Monitoring) in Normal Mode
). The multi-parameter
displays only error or alarm status messages.
The following hardware is needed for Philips VueLink™* communication:
•
Philips VOI B module (Philips VueLink™* P/N M1032A #A05, driver code TU1AA)
Pin #
Signal name
Pin #
Signal name
1
Data carrier detect
6
Data set ready
2
Receive data
7
Request to send
3
Transmit data
8
Clear to send
4
Data terminal ready
9
Ring indicator
5
Ground
Содержание INVOS 5100C
Страница 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Страница 12: ...Page Left Intentionally Blank x Operator s Manual...
Страница 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Страница 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Страница 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Страница 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Страница 142: ...Database Operational Mode and Area Under the Curve Page Left Intentionally Blank 8 24 Operator s Manual...
Страница 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Страница 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Страница 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Страница 177: ......