Guidance and Manufacturer’s Declaration—Electromagnetic Compatibility (EMC)
Operator's Manual
A-9
Table A-10.
Electromagnetic Immunity—Conducted and Radiated RF
Immunity test
IEC/EN 60601-1-2 test
level
Compliance level
Electromagnetic environment guid-
ance
Frequency of transmitter
Equation for separation distance (
d
)
Conducted RF
IEC/EN 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
150 kHz to 80 MHz
150 kHz to 80 MHz
Radiated RF
IEC/EN 61000-4-3
3 V/m
80 MHz to 800 MHz
3 V/m
80 MHz to 800MHz
80 MHz to 800 MHz
3 V/m
800 MHz to 2.5 GHz
3 V/m
80 MHz to 2.5 GHz
800 MHz to 2.5 GHz
P
is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and
d
is the
separation distance in meters (m).
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the mea-
sured field strength in the location in which the INVOS™ system is used exceeds the applicable RF compliance level above, the
system should be observed to verify normal operation. If abnormal performance is observed, additional measures may be nec-
essary, such as reorienting or relocating the INVOS™ system.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Interference may occur in the vicinity of equipment marked with the following symbol:
Table A-11.
Recommended Separation Distances
Rated maximum output power
(
P
) of transmitter in watts (W)
Separation distance (
d
) in meters
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12 0.12 0.23
0.1
0.38
0.38
0.73
1.00
1.20
1.20
2.30
10.00
3.80
3.80
7.30
100.00
12.00
12.00
23.00
For transmitters rated at a maximum output power not listed above, estimate the separation distance (
d
) using the equation
in the corresponding column, where
P
is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
d
1.2
P
=
d
1.2
P
=
d
2.3
P
=
d
1.2
·
P
=
d
1.2
P
=
d
2.3
P
=
Содержание INVOS 5100C
Страница 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Страница 12: ...Page Left Intentionally Blank x Operator s Manual...
Страница 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Страница 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Страница 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Страница 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Страница 142: ...Database Operational Mode and Area Under the Curve Page Left Intentionally Blank 8 24 Operator s Manual...
Страница 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Страница 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Страница 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Страница 177: ......