Maintenance
9-2
Operator's Manual
Caution:
Do not autoclave the INVOS™ system.
Caution:
Do not gas sterilize the INVOS™ system.
1.
Disconnect the AC power from the INVOS™ monitor. Turn the system off by selecting the green ON/
OFF
key.
2.
Clean the outside surface of the enclosure with a cloth dampened with Isopropyl alcohol–70% or a
mild soap and water solution.
3.
Clean the faceplate and screen with a clean, soft cloth and isopropyl alcohol or a commercial glass
cleaner. Do not use acetone or abrasives.
4.
If necessary, the INVOS™ system and cables can be wiped clean with commercial germicidal agents.
5.
Allow the INVOS™ system monitor to completely dry before reconnecting the AC power.
9.5
Battery Recharge
WARNING:
The backup battery is not user replaceable.
The battery charges itself while the INVOS™ system is connected to AC power. It acts as a backup
for data storage, power outages, and patient transport. The battery has an approximate 20-
minute operation capacity. Recharge time from 50% discharge to full charge is approximately 20
hours and 100% discharge to full charge is approximately 24 hours.
Note:
Upon initial unpacking the battery may be partially discharged. Before attempting battery operation, the
battery must be charged by connecting the INVOS™ system monitor to AC power in either operational or
nonoperational mode. A charge time of 24 hours is required to recharge a fully depleted battery. The blue
LED on the key panel illuminates to indicate the power is connected and the battery is charging.
Note:
The backup battery is not user replaceable. Any attempt to access the battery by the user will void the
product warranty.
Содержание INVOS 5100C
Страница 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Страница 12: ...Page Left Intentionally Blank x Operator s Manual...
Страница 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Страница 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Страница 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Страница 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Страница 142: ...Database Operational Mode and Area Under the Curve Page Left Intentionally Blank 8 24 Operator s Manual...
Страница 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Страница 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Страница 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Страница 177: ......